Mindray BS 200E
BS-200E
Chemistry Analyzer
Technical Specifications
System Function:
Automatic, Discrete, Random Access
STAT sample priority
Throughput: Constant 200 tests/hour (without ISE), up to
330 tests/hour with ISE
Principles: Absorbance photometry, Turbidimetry,
Ion Selective Electrode technology
Methodology: End-point, Fixed-time, Kinetic, optional ISE
Single/Dual reagent chemistries,
monochromatic/bichromatic
Linear/non-linear multi-point calibration
Programming: Open system with user defned profles
and chemistry calculation
System pack reagents ready to use
Reagent/Sample Handling:
Reagent/Sample tray:
40 reagent positions, 40 sample positions
in cooling compartment (2~12C)
Reagent volume:
R1: 10~350μl, step by 1 μl
R2: 10~200μl, step by 1 μl
Sample volume: 2~45μl, step by 0.1 μl
Reagent/Sample probe:
Liquid level detection and tracking, vertical &
horizontal collision protection and inventory
checking
Probe cleaning: Automatic washing of interior and exterior
Carry-over < 0.1%
Automatic sample dilution:
Pre-dilution and post-dilution
Dilution ratio up to 1: 200
Internal Bar Code Reader (optional):
Used for sample and reagent scan
Applicable to various bar code systems such as
Codabar, ITF (Interleaved
Two of Five), code128, code39,
UPC/EAN, Code93
Bi-directional interface LIS transmission
Mindray BS 200E
ISE Module (optional):
Measure K + , Na + , Cl
Optical System:
Light Source: Halogen-tungsten lamp
Photometer: Grating system, reversed optics
Wavelength: 12 wavelengths, 340nm, 380nm, 412nm, 450nm,
505nm, 546nm, 570nm, 605nm, 660nm, 700nm,
740nm and 800nm
Absorption range: 0~3.3Abs (10mm conversion)
Resolution: 0.0001Abs
Reaction System:
Reaction rotor: Rotating tray, containing 80 cuvettes
Cuvette: Reusable, optical length 5mm
Reaction volume: 150~500μl
Reaction temperature: 37C
Temperature fuctuation: ±0.1C
Mixing System: Standalone mixing bar
Cuvette Washing: 8-step washing station with pre-heated
detergent and water
Control and Calibration:
Calibration mode: Linear (one-point, two-point and multi-point),
Logit-Log 4P, Logit-Log 5P, Spline,
Exponential, Polynomial, Parabola
Control software: Westgard multi-rule, Cumulative sum
check, Twin plot, L-J Chart
Operation Unit:
Operation system: Windows XP Professional/Home SP2 or above
Windows 7
Interface: RS-232
Working Conditions:
Power Supply: AC 200~240V, 50/60Hz, ≤1500VA or
AC 100~130V, 50/60Hz, ≤1500VA
Temperature: 15-30C for operation
Humidity: 35-85% RH
Dimension: 860mm (W) x700mm (D) x625mm (H)
Weight: 130 Kg
MSL TEAM picture