BC-6800 Auto Hematology Analyzer Technical Specifications:
Principles
SF Cube cell analysis technology for WBC, 5-Part diff, NRBC, RET and PLT-O
Focusing Flow-DC method for RBC and PLT
Cyanide free hemoglobin measurement
Parameters
37 reportable parameters (whole blood): WBC, Lym%, Mon%, Neu%, Bas%, Eos%, IMG%, Lym#, Mon#, Neu#, Eos#, Bas#, IMG#; RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, RET%, RET#, IRF, LFR, MFR, HFR, RHE, NRBC#, NRBC%; PLT, MPV, PDW, PCT, P-LCR, P-LCC, IPF
17 research parameters (whole blood): HFC#, HFC%, WBC-R, WBC-D, WBC-B, WBC-N, RBC-O, PLT-O, PLT-I, PDW-SD, InR#, InR‰, Micro%, Micro#, Macro%, Macro#, MRV
7 reportable parameters (body fluid): WBC-BF, TC-BF#, MN#, MN%, PMN#, PMN%, RBC-BF
7 research parameters (body fluid): Eos-BF#, Eos-BF%, Neu-BF#, Neu-BF%, HF-BF#, HF-BF%, RBC-BF
2 histograms for RBC and PLT
3 scattergrams (3D) for DIFF, NRBC and RET
6 scattergrams (2D) for DIFF, BASO, NRBC, RET, RET-EXT, PLT-O
Mindray BC 6800 automated hematology analyzer Sample Volume
Predilute mode (capillary blood), Open vial 40μL
Manual mode (whole blood), Open vial 150μL
Autoloader mode (whole blood), Closed vial 200μL
Manual mode (body fluid), Open vial 150μL
Throughput
Up to 125 samples per hour (CBC+DIFF)
Up to 90 samples per hour (CBC+DIFF+RET)
Up to 40 samples per hour (body fluid)
Loading capacity
Up to 100 sample tubes
Mode
CBC, CBC+DIFF, CBC+RET, CBC+NRBC, CBC+DIFF+RET,
CBC+DIFF+NRBC, CBC+DIFF+RET+NRBC, RET
Data storage capacity
Up to 100,000 patient results including all numeric and graphical
information
Printout
Various printout formats and user-defined formats available
Operating environment
Temperature: 15oC~32oC
Humidity: 30%~85%