Mobile blood Dialysis Machine MSLHM01
MSL blood Dialysis Machine MSLHM01 Main performance of this device:
1. Self-checking function;
2. Carbonate dialysis;
3. Double needle dialysis;
4. Liquid level detector;
5. Bubble detector;
6. Blood leakage detector;
7. Temperature and electric conductivity monitoring;
8. Arterial pressure, venous pressure and transmembrane pressure monitoring;
9. Rolling blood pump;
10. Heparin pump;
11. Dehydration amount controlled by capacity;
12. Automatic disinfection cleaning program;
13. Stand-by power of blood pump in case of power failure;
14. Information display function of display screen.
*This device is designed and manufactured in strict accordance with the
Requirements of national laws and regulations. The reference standards of design and manufacture are following:
GB9706.1-2007; GB9706.2-2003; and YY0054-2003.
Cheap Mobile blood Dialysis Machine MSLHM01 Technical Parameter
Size &Weight:
Size: 380mmx400x1380mm (L*W*H);
Area: 500*520mm;
Weight: 88KG;
Power supply:
AC220V, 50Hz/60Hz, 10A
Input power: 1500W
Back-up battery: 30 minutes
Best Mobile blood Dialysis Machine MSLHM01 Working Condition:
Water input pressure: 0.1Mpa~0.6Mpa, 15P.S.I~60P.S.I
Water input temperature: 5℃~30℃
Working environment temperature: 10℃~30℃ at relative humidity ≦70%
UF rate:
Flow range: 0ml/h~4000ml/h
Resolution ratio: 1ml
Precision: ±30ml/h
Blood pump &Spare pump:
Blood pump flow range: 30ml/min~600ml/min (diameter: 8mm or 6mm)
Spare pump flow range: 30ml/min~600ml/min (diameter 8mm or 6mm)
Resolution ratio: 0.1ml
Precision: ±10ml or 10% of reading
Heparin pump:
Syringe size: 20, 30, 50ml
Flow range: 0ml/h~10ml/h
Resolution ratio: 0.1ml
Precision: ±5%
Monitoring system &Alarm set up:
Venous pressure: -180mmHg ~ +600mmHg, ±10mmHg
Arterial pressure: -380mmHg ~ +400mmHg, ±10mmHg
TMP pressure: -180mmHg ~ +600mmHg, ±20mmHg
Dialysate temperature: preset range 34.0℃~39.0℃
Blood leak detection: Photo chromic alarm when erythrocyte specific volume is 0.32±0.02 or blood leak volume is equal or more than 1ml per liter of dialysate.
Bubble detection: Ultrasonic, Alarm when a single air bubble volume is more than 200μl at 200ml/min blood flow
Conductivity: Acoustic-optic
Use Mobile blood Dialysis Machine MSLHM01 Disinfection/Sanitize:
1. Heat disinfection:
Time: 30 minutes; Temperature: about 80℃, at flow rate 500ml/min;
2. Chemical disinfection:
Time: 30minutes, Temperature: about 36℃~50℃, at flow rate 500ml/min;
3. Chemical disinfection with heat:
Time: 45minutes, Temperature: about 36℃~80℃, at flow rate 50ml/min;
4. Rinse:
Time: 10minutes, Temperature: about 37℃, at flow rate 800ml/min;
Storage Environment:
Storage temperature should be between 5℃~40℃, at relative humidity ≦80%