Technical Specifications
.System Function
Automatic, Random Access
STAT sample priority, available for all sample position
Barcode Scanner for sample and reagent
Bi-direction LIS interface (HL7)
Throughput:
360 tests/hour (single and dual reagent)
Up to 540 tests/hour with ISE (K+, Na+, CI-)
Principle: Colorimetric, Turbidimetric
Methodology:
End-point, Two-point, Kinetic Discrepant Value
Single/Dual/Triple/Quadruple reagent chemistries
Monochromatic/Bichromatic
Programming: Open system with user defined profiles and calculation assays
Probe cleaning:
Automatic washing both interior and exterior, carryover <0.1%
Pre-heating for reagent
Auto dilution:3-150 times
Built-in reagent/sample bar code reader (optional)
.Reaction System
Reaction positions: 100 cuvettes
Cuvette:Optical length 5mm
Reaction volume: 150.500u1
Reaction temperature:37'C,10.1℃
2 mixing probes
8 steps auto-washing system
.Sample/Reagent Handling
2 reagent probes, 1 sample probe
Sample tray:100 positions
Reagent tray: 80 refrigerated positions
Refrigeration temperature:2一100C
Reagent volume:20、350 NL, 1 pL adjustable
RI:150·350 pL, 1 pL adjustable
R2:20·250 NL,1 pL adjustable
Support R3 and R4
Sample volume: 2、45 pL, 0.1 pL adjustable
Sample and Reagent probe: Collision protection, liquid level detection and inventory checking
.Measuring and Optical System
Light Source:Halogen tungsten lamp
Photometer:Grating photometry
Wavelengths: 340nm,405nm, 450nm, 505nm, 540nm, 570nm, 600nm, 635nm,670nm,700nm,760nm,795nm
Absorption range:0~2.500Abs
Resolution:0.0001 Abs
.Working Condition
Power supply:100·240 VAC 50/60 Hz, 1500w
Temperature: 100C、300C
Humidity:40%~85%
Water consumption: s20 Uhour
Dimension: 1150mm(L)0726.5mm(W)01150mm(H)
.360 tests/hour, up to 540 tests/hour with ISE(optional)
.Grating system with 12 wavelengths
.Built-in bar code reader (optional)
.8 steps washing system
.Bi-direction LIS interface