No. |
Technical parameters |
1 |
X-ray generators |
1.1 |
Radiography & Fluoroscopy |
1.2 |
Power voltage: AC380V±38V |
1.3 |
Supply frequency: 50Hz/60Hz±1Hz |
1.4 |
Power: 50kW |
1.5 |
Max output power: 50kW |
1.6 |
Operating frequency ≥100kHz |
1.7 |
Tube current range |
Radiography:10mA~630mA |
Fluoroscopy:0.2mA~5.0mA |
1.8 |
Tube voltage range: |
Radiography:40~150kV, adjust lenght 1kV |
Fluoroscopy:40~125kV, adjust lenght 1kV |
1.9 |
Range of exposure duration:0.002s~6.3s (2ms~6300ms) |
1.1 |
current time:0.1mAs~630mAs |
2 |
Operations Console |
2.1 |
Dimension :1400mm×400mm×115mm |
3 |
X-ray Pipe sleeve |
3.1 |
Max tube voltage:150kV |
3.2 |
Thermal capacity:930 kJ(1300kHU) |
3.3 |
Inherent filtration:0.7mm AL/75kV |
4 |
X-ray tube |
4.1 |
Max tube voltage: 150 kV |
4.2 |
Max Filament current: 5.3A (Big) /5.4A (small) |
4.3 |
Bifocal:0.6mm(small focus)/1.2mm(big focus) |
4.4 |
Anode rated power:32KW(small focus);77KW(big focus) |
4.5 |
Target angle:12° |
4.6 |
Anode rotation speed:2850 rotate/minute |
4.7 |
Anode heat capacity:212kJ(300kHU) |
5 |
Beam limiter |
5.1 |
power supply: 24VAC 4.5A |
5.2 |
Inherent filtration:1.0mmAL/70kV |
5.3 |
Additional filtration:0/0.5/1.0/1.5mmAL/70kV(Switch inside the machine) |
6 |
Diagnostic bed power control cabinet |
7 |
Photography bed |
7.1 |
voltage:SimplexAC220V±22V |
7.2 |
Power frequency:50Hz±1 Hz |
7.3 |
input power:≤1kVA |
7.4 |
Panel bed size:Length × width × height=2100×775×635mm |
7.5 |
Bed mobile performance:Bed lateral movement range: ≥250mm |
Bed longitudinal movement range: ≥900mm |
8 |
X-ray tube assembly |
8.1 |
Maximum tube voltage:150kV |
8.2 |
Heat capacity:1100kJ(1540kHU) |
8.3 |
Inherent filtration:1.0mm Al |
8.4 |
Total filtration:2.5mm AL |
9 |
X-ray tube |
9.1 |
Operating voltage: 40~150kV |
9.2 |
The maximum filament current: 5.5A |
9.3 |
Bifocal:1.0mm(Small focus)/2.0mm(Big focus) |
9.4 |
Anode input power:30KW(Small focus);50KW(Big focus) |
9.5 |
Target surface angle:17..5° |
9.6 |
Anode rotation speed:2800 rev / min |
9.7 |
Anode heat capacity:140kJ(190kHU) |
10 |
Beam limiter |
10.1 |
power supply: 24VAC 4.5A |
10.2 |
Inherent filtration:1.0mmAL/70kV |
10.3 |
Additional filtration: 0/0.5/1.0/1.5mmAL/70kV ( switches inside the machine) |
11 |
Vertical type racks |
11.1 |
Vertical stroke:the distance between the top edge of film magazine and the ground should not be longer than 800mm, the stroke should no less than 1000mm |
11.2 |
Film size:the maximum cassette should not be bigger than 35.6cm×43.2cm |
11.3 |
aluminium base grid) convergence distance f0=180cm, grid ratio r=10, grid density N=40lines/cm |
12 |
Remote controlled diagnostic bed |
12.1 |
Supply voltage:single phase AC220V±22V |
12.2 |
Supply frequency: 50Hz±1 Hz |
12.3 |
Input power: ≤1kVA |
12.4 |
Bed panel dimension: 2000×750×835mm |
12.5 |
Bed surface horizontal movement range: 240mm±10mm |
12.6 |
Slewing area of the lathe body: 90°~ 0°~ -15° |
13 |
Image processor |
13.1 |
Power supply: AC100~240V 50Hz |
13.2 |
Rated power: 15W |
14 |
Monochrome monitor |
14.1 |
Power supply: AC100~240V 50-60Hz 0.5A |