MSLGA28 is an economic type of Anesthesia Machine.
The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose one kind of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
Reliable pneumatic transmission system.
Economic and proper configuration of technologies, optional as required by the Customer.
Modularized design and upgrading. Upgrading is available as required by the Customer.
Part of Anesthesia
■Main body High-strength engineering plastic rack, light,
beautiful and corrosion resistant
■Scope of application adult, child
■Gas source O2, N2O: 0.28~0.6MPa
■Flow meter O2: 0.05~1.0L/min 1.1~10L/min
N2O: 0.05~1.0L/min 1.1~10L/min
■Flow rate of rapid oxygen supply 25~75L/min
■Low oxygen pressure alarm There will be sound alarm when the oxygen
pressure < 0.2MPa
■Vaporizer it has the function of automatic compensation based on pressure,temperature,and flow rate. The regulation range of evaporator concentration is 0~5 vol%.
Among Halothane,Enflurane,Isoflurane and Sevoflurane,one can be chosen for application as required by the customer.
■Respiratory circuit working mode: all-close, semi-close, semi-open
APL:≤12.5 kPa
■Respiratory bellows bellows for adults, bellows for children
tidal volume range:0~1500 ml
Part of Ventilator
■Display mode High-definition 5.7” LCD screen display
■Ventilation mode IPPV, SIPPV, SIMV, MANUAL
■Ventilation parameters Tidal volume 50~1500ml
Rate 2~99 bpm
SIMV 2~20 bpm
I:E 2:1~1:8
Inspiratory trigger pressure -10~20cmH2O
PEEP 2~20cmH2O
Pressure limitation 10~60cmH2O
SIGH 1.5 times the inspiratory time/100bpm
■Parameters for ventilation monitoring tidal volume, minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, pressure - time waveform, flow - time waveform, PEEP, inspiratory trigger pressure,(average pressure, oxygen concentration, battery capacity)
■Security Alarm System
Oxygen concentration alarm upper limit setting range (21%~99%)
low limit setting range (18%~80%)
Airway pressure alarm upper limit setting range 5~60 cmH2O
low limit setting range 0~50 cmH2O
Minute ventilation volume alarm upper limit setting range 1~20 L/min
low limit setting range 0~19 L/min
Tidal volume upper limit setting range 100~1500ml
low limit setting range 0~1400ml
Sustained high-pressure alarm it will give alarm when stress have consistently
been higher than 2.5 cmH2O
(Suffocation alarm, Power alarm, Intubation off alarm, battery voltage alarm)
■Power AC 220V 50Hz (UPS, Storage battery)
■(Mechanical Arms) ■(End-tidal CO2 monitoring)
■(Specific interface of waste anesthesia gas)
*Items within brackets are optional.