Colony counter digital semi-automatic bacteria inspection MSLZJ017
MSLZJ017 colony counter is a digital display-style semi-automatic bacteria inspection instrument.
Counter probe pen, counting pool parts. Counter using CMOS integrated circuit design and
manufacturing. Black depth background count tanks, saving circular fluorescent side-illumination
contrast clear colonies. Rules of order in accordance with the bacteria count test instrument display
is designed as a three-digit, more than 300 the number of colonies growing in a petri dish, the
test sample should be diluted redo to ensure the accuracy of the count. This instrument can reduce
the labor intensity of the laboratory personnel, and improve work efficiency and quality of work.
The products are widely used in food, beverages, pharmaceuticals, biological products, health supplies,
drinking water, industrial waste water, the number of bacteria in clinical specimens testing. Essential
equipment for all levels of health and epidemic prevention station, environmental monitoring stations,
food hygiene supervision and inspection of the hospital, the biological products of the Institute for Drug
Control, food factory, daily chemical factory and tertiary institutions, research units laboratory.
Semi-automatic bacteria inspection MSLZJ017 main parameters
• Counter Capacity: 0 to 999
• Light source lamp power: 16W
• Total power consumption: <20W
• Power supply voltage: 220V ± 10%, 50Hz
• Volume: 280 × 230 × 90
• Weight: 1.4kg
Digital semi-automatic bacteria inspection MSLZJ017 Usage
a. power toggle switch count pool lights, the display shows "001" the probe pen plug the jack into the
instrument, press the "reset" key to zero the instrument into the working state.
b. placed in a petri dish be seized.
c. exploration pen in a petri dish bottom face of all the colonies one by one point. Each point one
should hear a "beep" sound to illustrate effective otherwise, should be the focus. In this case, the point
to the colonies was marked with color, display digital automatic accumulation.
d with a magnifying glass, double-check to confirm Points exhaustive count already finished.
E figures within the display screen is the number of colonies of the dish.
f. records digital Remove the dish. Press the "reset" button, the display into the initial state.
Precautions
a. instrument should be placed on a flat solid surface.
points colonies exploration pen not to be too inclined to gently point under bouncing sense,
the number is entered.
c. instrument should be moisture-proof, anti-rattling against direct sunlight, anti-acid erosion,
should be added after the dust cover.
attention to the prevention of the bacterial cultures pollution counting pool.
e instrument and probe pen are not disassemble. Found fault, experienced technical staff overhaul.
MSL TEAM picture
MSL Certificate