Overall performance
Equipment type Fully automatic ,discrete, STAT priority
Throughput (Theoretical max) 600T/H for photometry tests only
600T/H for only ISE tests
Test principle Colorimetry, turbidimetry, ISE
Analysis method End-point, kinetics,fixed-time,etc.support single/double wavelength
and 1-2 reagents item, linear and non-linear calibration
Simultaneous analysis item 88 colorimetric items and 3 ISE items (K, Na, Cl optional)
Packaging detail:Standard export package Delivery detail:within 7-10 workdays after receipt of payment |
Rear spectrophotometry of holographic concave flat field grating, Cluster - optical path,
realizing supermicro volume detection of reaction solution
60 nm polished probe, liquid level detection, clot detection and collision protection
Automatic rinsing for reaction cuvettes
Sample rerun and dilution function
English graphic operation software, supports various formats report printing supports
User defined combined batch input of patient information, applicable to fast setting for special hospital, blood station and physical examination
1.Overall performance
Equipment type Fully automatic ,discrete, STAT priority
Throughput (Theoretical max) 600T/H for photometry tests only
600T/H for only ISE tests
Test principle Colorimetry, turbidimetry, ISE
Analysis method End-point, kinetics,fixed-time,etc.support single/double wavelength
and 1-2 reagents item, linear and non-linear calibration
Simultaneous analysis item 88 colorimetric items and 3 ISE items (K, Na, Cl optional)
2.Sample unit
Sample position 115 sample positions, including 50 routine sample positions, 20 STAT
positions, 34 calibration positions, 8 QC positions, 3 detergent
positions. Continuously cooling at calibration and QC positions to keep
5-15℃ within 24 hours
Sample cuvettes specification Standard cup, original blood tube, multi-specification tube available(Φ12~16) mm ×(25~100)mm
Sample barcode I2of5、code128、code39、UPC/EAN、code93
Sample volume 2μl~35μl,0.1μl stepping
Sampling technology liquid level detection, clot detection and collision protection
Sample probe carrying rate automatic warm water cleaning. Carrying rate≤0.1%
Automatic sample dilution 3~170 times
3.Reagent unit
Reagent probe 2, with the function of liquid level detection and collision
Reagent volume 20μl~350μl, 1μl stepping
Reagent position double reagent disk, 45X2, continuous 5-15℃ cooling within
24 hours, loading 70ml and 20ml standard reagent kit
Reagent barcode code128
Reagent probe carrying rate automatic warm water cleaning. Carrying rate≤0.1%
4.Reaction unit
Reaction cuvette 120 positions optical plastic cup, optical diameter is 6 mm
Total volume of reaction liquid 150μl~450μl
Reaction temperature 37℃,±0.1℃
Reaction disk constant temperature circulating water
Mixing needle 2, after adding reagent, blending immediately
Reaction cuvette cleaning 7 stops 11 steps by warm water rinsing
Wastewater treatment two diffluence for high and low concentration waste water. With the function of waste liquid level alarming
5.Optical system
Light source 20W/12V halogen lamps.
Monochromator Grating photometry
Photoelectron road Rear spectrophotometry
Wavelength 340nm、380nm、405nm、450nm、480nm、505nm、546nm、
570nm、600nm、660nm、700nm、800nm
Detector photodiode LED array
OD linear range 0~3.3Abs
6.Operating system
PC operating system Windows XP
Analysis control software English version graphical operating software
Main function of software automatic calibration, automatic barcode scanning, item compounding test, reagent info management, serum index, whole reaction process monitoring, dirty cup memory evading, prevent cross-contamination procedure, patient information memory and association input, automatic report audit, data multiple parameter query, report format statistic and printing, reference range classification, alarming information classification, user level classification, automatic dormancy and wake, real-time online help.
Report printing report formats support the user-defined mode, QC and state information etc
PC configuration CPU ≥2.2GHz(dual-core processor); Memory ≥1G; Harddisk ≥ 160G;
17 inch LCD display; Stylus, inkjet or laser printer (optional)
System connection TCP/IP network connection, standard RS-232C
7.Others
Volume 1060mm×790mm×1150mm
Weight 300Kg
Power supply voltage AC 220V / 230V, 50 / 60Hz, power 2KVA
Client Usage Photos of Auto-Chemistry Analyzer CS-600
Please contact us if you want to know more details.
Medial & Video of Auto-Chemistry Analyzer CS-600
Please contact us if you want to know more details.
Previous:Buy High Quality Auto-Chemistry Analysers CS-1200 from Medsinglong
Next:Professional Full-Biochemistry,Auto-Chemistry Analysers CS-400
Address:85th, Baiyun Road, Shiqiao Street, Panyu District, GuangZhou City Sitemap
Email:Cindy@medmsl.com / cindy@medicalequipment-msl.com Mob:+86 13168310539
Connect us in Whatsapp or Facebook Messenger
Messenger