Brief Introduction
● Two control modes: Air mode and Baby mode controlled by micro-computer;
● Set temperature, Air temperature, Baby temperature are displayed separately;
● Passive humidification;
● Inclination of bassinet is adjustable;
● Multiple failure alarm indications;
● RS-232 connector for data output;
● Oxygen input interface;
● MSLBI16 is configured with fixed Neonate Bilirubin Phototherapy Equipment model MSLHD01 with blue fluorescence lamp,or MSLHD02 with blue LED as light source;
● Low noise DC motor.
Packaging detail:Standard export package Delivery detail:within 7-10 workdays after receipt of payment |
MSLBI16 Infant incubator | neonatal icu
Brief Introduction
● Two control modes: Air mode and Baby mode controlled by micro-computer;
● Set temperature, Air temperature, Baby temperature are displayed separately;
● Passive humidification;
● Inclination of bassinet is adjustable;
● Multiple failure alarm indications;
● RS-232 connector for data output;
● Oxygen input interface;
● MSLBI16 is configured with fixed Neonate Bilirubin Phototherapy Equipment model MSLHD01 with blue fluorescence lamp,or MSLHD02 with blue LED as light source;
● Low noise DC motor.
MSLBI16 Infant incubator | neonatal icu
Specification
Power requirement: AC220V-230V/50Hz or AC110-120V/50-60Hz, 850VA
Control Modes: Air mode and Baby mode controlled by micro-computer
Control range of air temperature: 25℃-37℃(override mode: 37℃-39℃ is optional)
Control range of baby temperature: 34℃-37℃(override mode: 37℃-38℃ is optional)
Accuracy of skin temperature sensor: ±0.2℃
The temperature fluctuation range between incubator temperature and average incubator temperature (Under stable temperature condition): ≤0.5℃
Uniformity of temperature: ≤0.8℃ (mattress is on the horizontal position)
Inclination of the bassinet: ±12°
Noise inside hood: ≤45dB(A)
Failure alarm: over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, system alarm and so on.
The average value of total irradiance for bilirubin Ebi max of photo-therapy equipment:≥640μW/cm 2 (MSLHD01)≥1300μW/cm 2 (MSLHD02)
Built-in timer range: 0~9999 hrs 59min
Accuracy of weight display: ±1% (equipped weighing system)
MSLBI16 Infant incubator | neonatal icu
Optional Configuration
">37℃" temperature override mode function, Vertical height adjustment stand(VHA stand), Weighing system, SpO 2 , Neonate Bilirubin Phototherapy Equipment which is MSLHD02 (Fixed model), Disposable skin temperature Sensor.
Environmental temperature
Environment temperature: 20℃~30℃(Set temperature should be 3℃ higher than the environment temperature)
Ambient air velocity: < 0.3m/s
MSL Certificate
MSL Medical cooperate with DHL,FEDEX,UPS,EMS,TNT,etc.International shipping company,make your goods arrive destination safely and quickly.
Welcome to medicalequipment-msl.com.
If you have any demand in medical equipment, please feel free to contact cindy@medicalequipment-msl.com.
Address:85th, Baiyun Road, Shiqiao Street, Panyu District, GuangZhou City Sitemap
Email:Cindy@medmsl.com / cindy@medicalequipment-msl.com Mob:+86 13168310539
Connect us in Whatsapp or Facebook Messenger
Messenger